MRI RF Head Coil Test Results These images were taken as part of independent lab tests at Hershey Medical Center on the first Doty quadrature rf Litzcage 3 T head coil.
(Fast spin echo; TE=80.4 ms; TR=3150 ms; NEX=2; 4 mm slice; 256 x 256; 25 x 25 x 25 cm3 FOV; 3:23 per slice) Doty Litzcage Bruker Birdcage
PP1 =47 W PP1 =67 W Peak power (total at coil) for 3 ms Gaussian 180° pulse.
T1 Weighted Head Images in Bruker 3T (Single spin echo; TE=12.1 ms; TR=456 ms; NEX=2; 4 mm slice; 256 x 256; 25 x 25 x 25 cm 3 FOV; 3:53 per slice) Doty Litzcage Bruker Birdcage
PP1 =47 W PP1 = 67 W Peak power (total at coil) for 3 ms Gaussian 180° pulse.
SNR in 50 mM NaCl Phantom (15.3 cm dia., 25 cm height; Fast Spin echo; TE=76 ms; TR=4023 ms; 5 mm slice; 256 x 256; axial: 20 x 20 cm2 FOV; sagittal: 20 x 32 FOV) SNR(w): signal from whole plane: SNR(c): central portion, 25% radius or height
|
|||